Analab Scientific Instruments Private Limited

Analab Scientific Instruments Private Limited

Email
Send Email

Karl Fischer Titrator

Analab Scientific Instruments Private Limited

Home  » Products  » Karl Fischer Titrator  » Karl Fischer Titrator, Model : µAquaCal10

Karl Fischer Titrator

The Karl Fischer Titrator, is ideally used in different industries for determining the quantity of water while manufacturing processes. These instruments are highly appreciated for their features like high accuracy, quick analysis preparation, small sample quantity required, auto drain facility and high durability. The range is idyllically used in chemical and pharmaceutical industrial manufacturing processes.

The Karl Fischer Titrator (Model : µAquaCal10), is ideally used in different industrial manufacturing processes for determining the amount of water. To suffice the water amount determining requirements of our clients, we offer these titrators with dual platinum electrode, moisture trap for reaction vessel, 250ml amber reservoir bottle, Teflon tubing for dispensing path, 250ml reaction vessel, Teflon paddle Teflon assembly, dispensing tube (02 nos) manufacturer test certificate, warranty card and operational and instruction manual.

Model : µAquaCal10

Supplied with 250ml Reaction Vessel, Dual Platinum Electrode, Teflon Paddle Teflon Assembly , Moisture Trap for Reaction Vessel & 250ml Amber Reservoir Bottle, Dispensing Tube (02 Nos) , Teflon Tubing for Dispensing Path, Operational & Instruction Manual, Manufacturer Test Certificate and Warranty Card.

Salient Feature

  • Metalic Epoxy Based Powder Coated Body
  • Built in Magnetic Stirrer with soft touch Key Speed Controller
  • Rugged Teflon Housing for holding Dual Platinum Electrode & Moisture Trap.
  • Automatic end point detection with Audio Alarm.
  • Auto Drain Facility for cleaning reaction vessel (Optional) OVER TITER Audio Alarm save KF Reagent Onsite Dispenser Calibration/Validation Facility.
  • MOC of Dispenser is SS/ABS/TEFLON for Corrosion Protection & maintenance Free Operation.
  • Flushing System for cleaning dispensing path.

Technical Specification

Method of Detection Conductometric
Backlit LCD Display 16 x 2 Line Alphanumeric
Range of Moisture Detection less than 20 PPM to 100%
Display of Reading in ml with minimum resolution 0.01ml.
Dispenser Motorized Dispensing for low & accurate measurement of reagent.
Electrode Dual Platinum Electrode
Stirring Time 00
End Point Time 1-99 Seconds , Users Adjustable
End Point Sensitivity of Moisture Detection Users Adjustable
Input Electrode Connector
Power supply 230 VAC ± 10% , 50 Hz

Optional

  • Auto Drain of Reaction Vessel

Analysis Report

Report Date & Time 12-FEB-2010 – 16:9
Company Name ANALAB (VADODARA)
Calibration Date & Time 11-FEB-2010 – 23.34
Calibration Method Water by Volume
Average Titer Factor 6.666666
Record Number 00
Analysis Date & Time 12-FEB-2010 – 16.00
Sample Type & Method Liquid by Volume
Volume of Sample (ml.)) 05.000
Density of Sample 0.789
Stirring Time (Min:Sec) 0
End Point Time (Seconds) 20
End Point (ml.) 01.21
Moisture % 000.2044
Moisture (PPM) 0002044.0
H2O (mg/ml) 003260.33


Yes! I am Interested