Analab Scientific Instruments Private Limited

Analab Scientific Instruments Private Limited

Email
Send Email

pH/mV/ºC/ORP Analyzer

Analab Scientific Instruments Private Limited

Home  » Products  » pH/mV/ºC/ORP Analyzer   » pH / mV / ºC Analyzer

pH / mV / ºC Analyzer

The hand held analyzer offers up to five point calibration and is supplied with combination glass pH electrode and standard buffer tablets. The product also comes with temperature probe, electrode stand, dust cover and operational and instructional manual.

Model : µpHCal100 (Three Point Calibration)

Supplied with Combination Glass pH Electrode (0-70ºC), Standard Buffer Tablets, Temperature Probe, Electrode Stand with Clamp, Dust Cover, DC Adaptor, Operational & Instruction Manual, Manufacturer Test Certificate and Warranty Card.

Sailent Features

  • Auto Buffer Recognition
  • Simultaneous Display of pH, mV & Temperature, Parallel & Serial Port for Printer & PC attachment
  • Report Formation as per GLP Compliance
  • More than 150 analysis data storage with corresponding
  • Calibration data with Batch Number, Real Date & Time, Operator Name, Sample Name, searching of data as per Batch Number Calibration Reminder Alarm (0-99 Hrs)
  • Results Hold / Un-Hold Facility etc.

Technical Specification

pH Range -1.000 to 15.000 pH
Selectable Resolution 0.1 / 0.01 / 0.001 pH
Relative Accuracy ± 0.002 pH
Range - mV / RmV ± 2000.0
ORP EH autocalibration to standard hydrogen electrode
Resolution 0.1
Relative Accuracy ± 0.2 mV or 0.05%, whichever is greater
Temperature Range 0 to 130.0ºC
Temperature Resolution 0.1ºC
Relative Accuracy ± 0.1 °C
Backlit LCD Display 20 x 2 line Alphanumeric
Auto Buffer Recognition 1.68, 4.01, 6.86 / 7.00, 9.18 / 10.01 & 12.45 pH @25ºC
Buffer Standard US / NIST & EURO
Power Supply DC Adaptor

Optional for µpHCal

  • RS-232 to USB interface with Cable & Driver 100 Software to Transfer Data From Instrument to PC

Error Message

  • Bad Buffer / Electrode Temperature Probe pH over range

Analysis Report

  • Company Name : ANALAB (VADODARA)
  • Analysis Date & Time : 13-NOV-2009 – 16:40
    pH = 04.291
    mV = +0160.2
    Temp. = 025.0

Analysis Report

  • Company Name : ANALAB (Vadodara)
  • REPORT PRINT DATE & TIME : 13-NOV-2009 – 16:55
  • CALIBRATION POINT : FIVE
  • Calibration Date & Time : 19-Jan-2010 – 15.27

Std. Buffer 4.01 Std. Buffer 7.00/6.86 Std. Buffer 10.01/9.18
pH = 04.291 pH = 07.185 pH = 09.589
mV = +0160.2 Offset = -0011.0 mV = -0153.2
Temperature = 025.0 Temperature = 025.0 Temperature = 025.0
% Slope S1 = 096.77 Buffer used : 07.00 % Slope S2 = 110.27

Std. Buffer 1.68 Std. Buffer 12.45
pH = 01.680 pH = 12.451
mV = +0314.7 Offset = -0011.0
Temperature =025.0 Temperature = 025.0
% Slope S1: 100.09 % Slope S4 = 099.93